9 conductivity, 1 general information, 2 measuring – Xylem Multi 3410 User Manual
Page 78: Conductivity 4.9.1, General information, Measuring

Operation
Multi 3410
78
ba75863e11
02/2011
4.9
Conductivity
4.9.1
General information
You can measure the following parameters:
z Conductivity
z Specific resistance
z Salinity
z Total dissolved solids (TDS)
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This enables interference-free measurements in
grounded measuring media.
Temperature
measurement
IDS sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
Preparatory activities
Perform the following preparatory activities when you want to measure:
4.9.2
Measuring
You can carry out conductivity measurements as follows:
1
Connect an IDS conductivity sensor to the meter.
The conductivity measuring window is displayed.
The Measuring cell and cell constant for the connected IDS
conductivity sensor are automatically taken over.
1
Perform the preparatory activities according to section 4.9.1.
2
Immerse the IDS conductivity sensor in the test sample.
22.09.2009 08:00