3 editing the measurement data storage – Xylem Multi 3410 User Manual
Page 92

Operation
Multi 3410
92
ba75863e11
02/2011
4.10.3 Editing the measurement data storage
The contents of the manual or automatic measurement data storage
can be shown on the display.
Each of the measurement data storages has a function to erase the
entire contents.
Editing the data storage
The storage is edited in the menu, Storage & config / Data storage. To
open the Storage & config menu, press the <MENU/ENTER
_
> key in
the measured value display.
Open the manual or automatic storage directly with the <RCL> or
<RCL
_
> key.
Note
The settings are explained here using the manual storage as an exam-
ple. The same settings and functions are available for the automatic
storage.
Settings
Menu item
Setting/
function
Description
Data storage /
Manual data storage /
Display
-
Displays all measurement
datasets page by page.
Further options:
z Scroll through the data-
sets with <
W><X>.
z Output the displayed
dataset to the interface
with <PRT>.
z Quit the display with
<ESC>.
Data storage /
Manual data storage /
Copy to USB flash
drive
-
Outputs all stored measure-
ment data to a connected
USB flash drive
Data storage /
Manual data storage /
Output to RS232/USB
-
Outputs all stored measure-
ment data to the interface
Data storage /
Manual data storage /
Erase
-
Erases the entire manual
measurement data storage.
Note:
All calibration data remains
stored when this action is
performed.