Xylem Multi 3410 User Manual

Page 52

Advertising
background image

Operation

Multi 3410

52

ba75863e11

02/2011

Continuing with two-

point calibration

9

Thoroughly rinse the IDS pH sensor with deionized water.

10

Immerse the IDS pH sensor in buffer solution 2.

11

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

S><T>.

12

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

13

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

14

If necessary, finish the calibration procedure as a two-point cal-
ibration with <M>.
The calibration record is displayed.

22.09.2009 08:00

Advertising