Xylem Multi 3410 User Manual
Page 48

Operation
Multi 3410
48
ba75863e11
02/2011
Sample record
Calibration evaluation
After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
Note
For pH IDS sensors, you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 4.6.10).
Multi 3410
Ser. no. 09250023
CALIBRATION
pH
Calibration date
31.07.2009 16:13:33
SenTix 940
Ser. no. B092500013
TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage
1
184.0 mV 24.0 °C
Voltage
2
3.0 mV
24.0 °C
Voltage 3
-177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...
Display
Calibration
record
Zero point
[mV]
Slope
[mV/pH]
+++
-15 ... +15
-60.5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the IDS sensor according to the
sensor operating manual
Error
Error
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to chapter
6 W