Atec Agilent-E8802A User Manual
Page 12

8
Table 6. Corrected System Performance With 7-16 Device Connector Type, 85038A Calibration
Kit
Applies to the E8801A, E8802A, and E8803A analyzer, 85038A (7-16, 50
Ω
) calibration kit, N6314A test port cable, and a full 2-port
calibration. Also applies to the following conditions:
•
IF bandwidth = 10 Hz
•
No averaging applied to data
•
Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
•
Isolation calibration not omitted
Specification (dB)
Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 to
6 GHz
6 to
9 GHz
a
Directivity
40
40
36
36
Source Match
37
37
34
34
Load Match
39
39
35
35
Reflection Tracking
±0.089
±0.089
±0.115
±0.115
Transmission
Tracking
±0.024
±0.033
±0.082
±0.103
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)