Atec Agilent-E8802A User Manual
Page 14

10
Table 9. Test Port Output Power
a
Description
Specification
Supplemental Information
Level Accuracy:
300 kHz to 6 GHz
±1.0 dB
Variation from 0 dBm in power range 0
6 GHz to 9 GHz
±2.0 dB
±1.5dB below 10 MHz
Level Linearity:
Variation from 0 dBm in power range 0
300 kHz to 9 GHz
±0.3 dB
-15 to +5 dBm
300 kHz to 1 MHz
±1.0 dB
+5 to +10 dBm
1 MHz to 6 GHz
±0.5 dB
+5 to +10 dBm
6 GHz to 9 GHz
±0.5 dB
+5 to +7 dBm
Range
b:
300 kHz to 6 GHz
-15 to +10 dBm
6 GHz to 9 GHz
-15 to +7 dBm
Range
b:
(Option 1E1):
300 kHz to 6 GHz
-85 to +10 dBm
6 GHz to 9 GHz
-85 to +7 dBm
Sweep Range
300 kHz to 6 GHz
25 dB
6 GHz to 9 GHz
22 dB
Level Resolution
0.01 dB
a
Source output performance on port 1 only. Port 2 output performance is typical.
b
Power to which the source can be set and phase lock is assured.