Corrected System Performance with Type-N Connectors
Table 2. Corrected System Performance With Type-N Device Connectors, 85032F Calibration
Kit
Applies to the E8801A, E8802A, and E8803A analyzer, 85032F (Type-N, 50
Ω
) calibration kit, N6314A test port cable, and a full 2-port
calibration. Also applies to the following conditions:
•
IF bandwidth = 10 Hz
•
No averaging applied to data
•
Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
•
Isolation calibration not omitted
Specification (dB)
Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6GHz to
9 GHz
Directivity
49
46
40
38
Source Match
41
40
36
35
Load Match
49
45
39
37
Reflection Tracking
±0.011
±0.021
±0.032
±0.054
Transmission
Tracking
±0.012
±0.020
±0.055
±0.083
Transmission Uncertainty (Specifications)