Atec Agilent-E8802A User Manual
Page 9

5
Table 3. Corrected System Performance With Type-N Device Connectors, 85092C Electronic
Calibration Module
Applies to the E8801A, E8802A, and E8803A analyzer, 85092C (Type-N, 50
Ω
) electronic calibration (ECal) module, N6314A test port cable,
and a full 2-port calibration. Also applies to the following conditions:
•
IF bandwidth = 10 Hz
•
No averaging applied to data
•
Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
•
Isolation calibration not omitted
Specification (dB)
Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 to
9 GHz
a
Directivity
52
54
52
47
Source Match
45
44
41
36
Load Match
47
47
44
39
Reflection Tracking
±0.040
±0.040
±0.060
±0.070
Transmission
Tracking
±0.039
±0.039
±0.068
±0.136
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)