Atec Agilent-E8802A User Manual
Page 13

9
Table 7. Uncorrected Instrument Performance
Specification (dB)
Description
300 kHz to
1 MHz
1MHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
9 GHz
Directivity
30
33
27
20
13
Source Match
18
18
16
11
8
Load Match
20
20
17
13.5
13
Reflection
Tracking
±1.5
±1.5
±1.5
±2.5
±3.0
Transmission
Tracking
±1.5
±1.5
±1.5
±2.5
±3.0
Test Port Output Characteristics (Source)
Table 8. Test Port Output Frequency
Description
Specification
Supplemental Information
Range:
E8801A
300 kHz to 3.0 GHz
E8802A
300 kHz to 6.0 GHz
E8803A
300 kHz to 9.0 GHz
Resolution:
1 Hz
Source Stability
±1 ppm, 0°C to 40 °C, typical
±1ppm/year maximum
Source Stability
(Option 1E5)
±0.05 ppm, 0° to 70 °C, typical
±0.1 ppm/year maximum
CW Accuracy
±3 ppm
CW Accuracy
(Option 1E5)
±1 ppm