Atec Agilent-E8802A User Manual
Page 17

13
Table 12. Test Port Input (Trace Noise)
Description
Specification
Supplemental Information
Trace Noise
a
Magnitude
1 kHz IF Bandwidth
< 0.002 dB rms
10 kHz IF Bandwidth
< 0.005 dB rms
Trace Noise
a
Phase
1 kHz IF Bandwidth
< 0.010° rms
10 kHz IF Bandwidth
< 0.035° rms
a
Trace noise is defined as a ratio measurement of a through or a full reflection, with the source set to 0 dBm.
Table 13. Test Port Input (Reference Level and Stability)
Description
Specification
Supplemental Information
Reference Level Magnitude
Range
±200 dB
Resolution
0.001 dB
Reference Level Phase
Range
±500°
Resolution
0.01°
Stability Magnitude
a
300 kHz to 3 GHz
0.02 dB/°C, typical
3 GHz to 6 GHz
0.04 dB/°C, typical
6 GHz to 9 GHz
0.06 dB/°C, typical
Stability Phase
a
300 kHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 9 GHz
0.2°/°C, typical
0.3°/°C, typical
0.6°/°C, typical
a
Stability is defined as a ratio measurement at the test port.