Atec Agilent-E8802A User Manual
Page 22

18
Table 19. Rear Panel Information (continued)
Description
Supplemental Information
External AM Input
Description
Input provides low-frequency AM modulation to test port output
signal, or shifts the test port output. Zero volts input gives the power
level set by the instrument, a positive voltage gives a higher level,
and a negative voltage gives a lower level.
Connector
BNC, female
Input Sensitivity
8 dB/V, typical
Bandwidth
1 kHz, typical
Input Impedance
1 k
Ω
, typical
External Detector Input
Description
Input from an external, negative polarity diode detector provides
ALC for a test port remote from instrument's front panel
Connector
BNC, female
Input Sensitivity
-500 mV yields approximately -3 dBm at detector's input, typical
Bandwidth
50 kHz, typical
Input Impedance
1 k
Ω
, nominal