Atec Agilent-E5071B User Manual
Page 12
Advertising

12
Table 1-12
Test port input (stability
1
)
Description Specification
Supplemental
information
Stability magnitude
3 MHz to 3 GHz
0.005 dB/° C
(at 23° C ±5° C, typical)
3 GHz to 6 GHz
0.01 dB/° C
(at 23° C ±5° C, typical)
6 GHz to 8.5 GHz
0.04 dB/°C
(at 23° C ±5° C, typical)
Stability phase
3 MHz to 3 GHz
0.1°/° C
(at 23° C ±5° C, typical)
3 GHz to 6 GHz
0.2°/° C
(at 23° C ±5° C, typical)
6 GHz to 8.5 GHz
0.8°/° C
(at 23° C ±5° C, typical)
Table 1-13 Test port input (dynamic accuracy)
Accuracy of the test port input power reading is relative to -10 dBm reference input power level.
Specification
Supplemental
information
Magnitude
Phase
1
Stability is defined as a ratio measurement at the test port.