Atec Agilent-E5071B User Manual
Page 8

8
Table 1-6
Uncorrected system performance (correction: off, system correction: on)
Description
Specification
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity
25 dB
20 dB
15 dB
Source match
25 dB
20 dB
15 dB
Load match
17 dB
12 dB
10 dB
Transmission tracking
±1.0 dB
±1.0 dB
±1.0 dB
Reflection tracking
±1.0 dB
±1.0 dB
±1.0 dB
Table 1-7
Test port output frequency
Description Specification
Supplemental
information
Range
E5070B
300 kHz to 3 GHz
E5071B
300 kHz to 8.5 GHz
Resolution 1
Hz
Source stability
Option E5070B/E5071B-UNQ
±5 ppm (5° C to 40° C, typical)
Option E5070B/E5071B-1E5
±0.05 ppm (23° C ± 5° C, typical)
±0.5 ppm/year (typical)
CW accuracy
Option E5070B/E5071B-UNQ
±5 ppm, 23° C ± 5° C
Option E5070B/E5071B-1E5
±1 ppm, 23° C ± 5° C
Uncorrected system performance
Test port output (source)