Table 1-3 Corrected system performance with type-N device connectors, 85092C electronic calibration module
Network analyzer: E5070B/E5071B, calibration module: 85092C (type-N, 50 Ω) electronic calibration (ECal) module, calibration: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with < 1°C deviation
from calibration temperature, isolation calibration not omitted
Description
Specification
(dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity 52 52 47
Source match
45
41
36
Load match
47
44
39
Reflection tracking
±0.040
±0.060
±0.070
Transmission tracking
±0.039
±0.069
±0.136
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)