Table 1-4 Corrected system performance with 3.5 mm device connector type, 85033E calibration kit
Network analyzer: E5070B/E5071B, calibration kit: 85033E (3.5 mm, 50 Ω), calibration: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with < 1°C deviation from
calibration temperature, isolation calibration not omitted
Description
Specification
(dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity 46 38 38
Source
match
43 37 36
Load
match
46 38 38
Reflection
tracking
±0.006 ±0.009 ±0.010
Transmission
tracking
±0.016 ±0.065 ±0.079
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)
Magnitude Phase