Table 1-5 Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration module
Network analyzer: E5070B/E5071B, calibration module: 85093C (3.5 mm, 50 Ω) electronic calibration (ECal) module, calibration: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C ± 5° C with < 1° C deviation from
calibration temperature, isolation calibration not omitted
Description
Specification
(dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity 52 51 47
Source match
44
39
34
Load match
47
44
40
Reflection tracking
±0.030
±0.050
±0.070
Transmission tracking
±0.039
±0.069
±0.117
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)
Magnitude Phase