Atec Agilent-E5071B User Manual
Page 4
Advertising
4
Table 1-2 Corrected system performance with type-N device connectors, 85032F calibration kit
Network analyzer: E5070B/E5071B, calibration kit: 85032F (type-N, 50 Ω), calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature = 23°C ±5°C with < 1°C deviation from calibration
temperature, isolation calibration not omitted
Description
Specification
(dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity 49 40 38
Source match
41
36
35
Load match
49
40
37
Reflection tracking
±0.011
±0.032
±0.054
Transmission tracking
±0.016
±0.062
±0.088
Transmission uncertainty (specification)
Magnitude Phase
Reflection uncertainty (specification)
Magnitude Phase
Advertising