Table 4-1
Corrected system performance with type-N 75 Ω device connectors, 85036E calibration kit
Network analyzer: E5070B/E5071B, calibration kit: 85036E (type-N 75 Ω), 50 Ω to 75 Ω adapters: 11852B, calibration: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C ± 5° C with < 1° C deviation from
calibration temperature, Isolation calibration not omitted
Description
Supplemental information (dB, typical)
3 MHz to 3 GHz
Directivity 37
Source match
33
Load match
37
Reflection tracking
±0.017
Transmission tracking
±0.021
Transmission uncertainty 3 MHz to 3 GHz (supplemental information, typical)
Magnitude Phase
Reflection uncertainty 3 MHz to 3 GHz (supplemental information, typical)
Magnitude Phase
Corrected system performance for 75 Ω measurements with 11852B
50 Ω to 75 Ω minimum-loss pads (supplemental information)