B&K Precision 9116 - Programming Manual User Manual

Page 19

Advertising
background image

readings.

Command syntax: FETCh:VOLTage?

Return parameter: <NRf>

Return parameter unit: V

MEASure[:SCALar]:CURRent[:DC]?

This command is used to read the actual current.

Command syntax: MEASure[:SCALar]:CURRent[:DC]?

Parameter: None

Return parameter: <NRf>

Return parameter unit: A

example: MEAS:CURR?

FETCh:CURRent?

This command is used to read the current which is in the sample cache. After sending the

command, the readings will be sent to the computer. This command does not affect the

instrument settings. This command does not trigger a measurement operation, and queries

only the latest available reading. Before reading the new reading, the command returns the old

readings.

Command syntax: FETCh:CURRent?

Return parameter: <NRf>

MEASure[:SCALar]:POWer[:DC]?

This command is used to enquire the actual output power.

Command syntax: MEASure[:SCALar]:POWer[:DC]?

Parameter: None

Return parameter: <NRf>

Return parameter unit: W

Example: MEAS:POWer?

FETCh:POWer?

This command is used to read the power which is in the sample cache. After sending the

command, the readings will be sent to the computer. This command does not affect the

instrument settings. This command does not trigger a measurement operation, and queries

only the latest available reading. Before reading the new reading, the command returns the old

readings.

Command syntax: FETCh:POWer?

Return parameter: <NRf>

Advertising