2 status subsystem, Status subsystem – B&K Precision 9116 - Programming Manual User Manual

Page 8

Advertising
background image

Query syntax: *STB?

Parameter: None

Returned parameter: <NR1>

*TRG

When power supply trigger source is a command from via BUS, this command will give a trigger

signal. And its function is the same as “TRIGger” command.

Query syntax: *TRG

Parameter: None

Returned parameter: None

*SAV

This command can save the current setups of power supply to specified memory. These setups

contain current setups, voltage setups, max voltage set, min voltage set, OVP set, OVP timer,

OPP set, voltage rise and fall time, analog interface and series/parallel setting.

The memory is divided into 10 groups, each contain 0~9 (10 total) setups. Up to 100 setups can

be saved in total.

Command syntax: *SAV<NRf>

Parameter: 0~9

*RCL

This command can recall the setups you saved previously from the specified memory location.

Command syntax: *RCL<NRf>

Parameter: 0~9

2.2 STATUS Subsystem

You can get the current status of the power supply by reading the operation status registers. The

power supply records the different status of the instrument through the four status register

group. The four status register groups are: status byte register, standard event register, query

status register and operation status register. Status byte register records the information of the

other status registers.

BIT Signal

Meaning

0

3

4

5

CAL

WTG

CV

CC

Operation status register

The power supply is calculating the new calibration parameter.

The power supply is waiting for the trigger information.

The power supply is in constant voltage status.

The power supply is in constant current status.

0

OV

Query status register

Over voltage