100

<decimal> - Voltage value. 0 – max. rated voltage

value are valid.

Example:

MEM:VSET2 4.132

Sets 4.132 V as voltage for the selected memory

location.

Query:

MEM:VSET2?

Return String:

4.132

Voltage is currently set to 4.132 V in the selected

memory location for channel 2.

Format:

<NRf>


MEMory:ISET

Description:

Configures/Queries the current value to store at the

selected memory location (use “MEM” command to

select memory location first)

Syntax:

MEMory:ISET<SP><decimal>

<decimal> - Current value. 0 – max. rated current

value are valid.

Example:

MEM:ISET 2.159

Sets 2.159 A as current for the selected memory

location.

Query:

MEM:ISET?

Return String:

2.159

Current is currently set to 2.159 A in the selected

memory location.

Format:

<NRf>


MEMory:ISET2

Note:

For dual channel models only.

Description:

Configures/Queries the second channel current value to

store at the selected memory location (use “MEM”

command to select memory location first)

Syntax:

MEMory:ISET2<SP><decimal>

<decimal> - Current value. 0 – max. rated current

value are valid.

Example:

MEM:ISET2 2.159

Sets 2.159 A as current for the selected memory

location.

Query:

MEM:ISET2?

Return String:

2.159

Current is currently set to 2.159 A in the selected

memory location for channel 2.