Voltage calibration, 6 ½ digit dmm – B&K Precision 9171-9174 - Manual User Manual
Page 175
Advertising

163
Voltage Calibration
- Use the setup diagram shown in Figure 33 below for voltage
calibration. Connect the rear output terminals of the power
supply to the DMM.
Figure 33- Voltage Calibration Setup
Follow the steps below to proceed with voltage calibration:
1. From the calibration menu, press to enter
VOLTAGE
CALIBRATION. The following screen will be shown.
2. Select the channel (for dual channel models), then press .
The calibration process will automatically start.
VOLTAGE CALIBRATION CH: 1
VL Set = X . XXXXV Read = XXXX
VML Set = XX . XXXXV Read = XXXXX
Enter
2
6 ½ Digit DMM
Advertising
This manual is related to the following products: