B&K Precision 9171-9174 - Manual User Manual
Page 73
Advertising

61
WARNING:
DO NOT output 10A or more with only one pair of
(+) and (-) terminals in the rear panel. Both (+)
terminals and both (-) terminals must all be
connected for applications requiring more than
10A output. Each terminal can only accept
maximum of 10A. Exceeding this may damage the
power supply.
+S + + - - - S
Rear Panel Output
Remove Shorting bars
+S + + - - - S
DUT
+ -
+S + - -S
Remove shorting bars
Front Panel Output
Rear Panel Output
Figure 19- Remote Sense Setup for 10A or Higher Current Output Operation
Advertising
This manual is related to the following products: