[SOURce:]CURRent:SLEW:NEGative
Channel Specific
This command sets the slew rate of the current for negative transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW:NEGative <NRf+>
Parameters see specifications
Unit A (amps per micro second)
Examples CURR:SLEW:NEG 0.0001
Query Syntax [SOURce:]CURRent:SLEW:NEGative?
Returned Parameters <NR3>
Related Commands CURR:SLEW
[SOURce:]CURRent:PROTection:STATe
Channel Specific
This command enables or disables the overcurrent protection feature.
Command Syntax [SOURce:]CURRent:PROTection:STATe <Bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples CURR:PROT:STAT 1
Query Syntax [SOURce:]CURRent:PROTection:STATe?
Returned Parameters <NR3>
Related Commands CURR:PROT
[SOURce:]CURRent:PROTection
Channel Specific
This command sets the soft current protection level. If the input current exceeds the soft current
protection level for the time specified by CURR:PROT:DEL, the input is turned off.
NOTE: Use CURR:PROT:DEL to prevent momentary current limit conditions caused by programmed
changes from tripping the overcurrent protection.
Command Syntax [SOURce:]CURRent:PROTection:LEVel <NRf+>
Parameters 0 through max. Rated current
Unit A (amps)