B&K Precision MDL Series - Programming Manual User Manual

Page 56

Advertising
background image

56

Channel Specific
These commands specify the transient level of the input current. The transient function switches

between level A and level B.

Command Syntax [SOURce:]CURRent:TRANsient:ALEVel <NRf+>

[SOURce:]CURRent:TRANsient:BLEVel <NRf+>

Parameters 0 through max. Rated current

Unit A (amps)

Examples CURR:TRAN:ALEV 5 CURR:TRAN:BLEV 0.5

Query Syntax [SOURce:]CURRent:TRANsient:ALEVel?

[SOURce:]CURRent:TRANsient:BLEVel

Returned Parameters <NR3>

Related Commands CURR:

[SOURce:]CURRent:TRANsient:AWIDth

[SOURce:]CURRent:TRANsient:BWIDth

Channel Specific
These commands specify the transient pulse width of the input current.

Command Syntax [SOURce:]CURRent:TRANsient:AWIDth <NRf+>

[SOURce:]CURRent:TRANsient:BWIDth <NRf+>

Parameters 1 to 65535 us

Unit S (seconds)

*RST Value 500uS

Examples CURR:TRAN:AWID 0.001 CURR:TRAN:BLEV 0.02

Query Syntax [SOURce:]CURRent:TRANsient:AWIDth?

[SOURce:]CURRent:TRANsient:BWIDth

Returned Parameters <NR3>

Related Commands CURR:

[SOURce:]CURRent:HIGH

[SOURce:]CURRent:LOW

Channel Specific

These commands set the high and low voltage level when the load is in constant current mode.

Command Syntax [SOURce:]CURRent:HIGH <NRf+>

[SOURce:]CURRent:LOW <NRf+>

Parameters see specifications for voltage range

Unit V (volts)

Examples CURR:HIGH 5