B&K Precision MDL Series - Programming Manual User Manual
Page 65

65
[SOURce:]POWer:PROTection
Channel Specific
This command sets the soft power protection level. If the input power exceeds the soft power
protection level for the time specified by POW:PROT:DEL, the input is turned off.
NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions caused by programmed
changes from tripping the overpower protection.
Command Syntax [SOURce:]POWer:PROTection[:LEVel] <NRf+>
Parameters 0 to max. rated power
Unit W (power)
*RST Value MAXimum
Examples POW:PROT 100
Query Syntax [SOURce:]POWer:PROTection[:LEVel]?
Returned Parameters <NR3>
Related Commands POW:PROT:DEL
[SOURce:]POWer:PROTection:DELay
Channel Specific
This command specifies the time that the input power can exceed the protection level before the input
is turned off.
Command Syntax [SOURce:]POWer:PROTection:DELay] <NRf+>
Parameters 0 to 60 seconds
Unit seconds
*RST Value 0
Examples POW:PROT:DEL 5
Query Syntax [SOURce:]POWer:PROTection:DELay?
Returned Parameters <NR1>
Related Commands POW:PROT
[SOURce:]POWer:CONFig
Channel Specific
This command sets the hard power protection level.