List commands – B&K Precision MDL Series - Programming Manual User Manual
Page 68

68
Command Syntax [SOURce:]IMPedance:HIGH <NRf+>
[SOURce:]IMPedance:LOW <NRf+>
Parameters 0 to max. rated voltage
Unit V (volts)
Examples IMP:HIGH 5
Query Syntax [SOURce:]IMPedance:HIGH?
[SOURce:]IMPedance:LOW?
Returned Parameters <NR3>
List Commands
List commands allow you to program complex sequences of input changes with rapid, precise timing,
and synchronized with trigger signals. Each function for which lists can be generated has a list of values
that specify the input at each list step.
[SOURce:]LIST:RANGe
Channel Specific
This command sets the current range for list mode. It also limits the maximum settable current.
Depending on the value, range will automatically switch between high or low depending on which
gives the best resolution and accuracy.
Command Syntax [SOURce:]LIST:RANGe <NRf>
Parameters 0 to max. rated current.
Unit None
Examples LIST:RANGE 30
Query Syntax [SOURce:]LIST:RANGe?
Returned Parameters <NR3>
Related Commands LIST:LEV
[SOURce:]LIST:COUNt
Channel Specific