B&K Precision MDL Series - Programming Manual User Manual

Page 62

Advertising
background image

62

[SOURce:]RESistance:HIGH
[SOURce:]RESistance:LOW

Channel Specific
This command is used to set the voltage high and low limit determined when the load is in constant

resistance mode.

Command Syntax [SOURce:]RESistance:HIGH <NRf+>

[SOURce:]RESistance:LOW <NRf+>

Parameters 0 to max. rated voltage

Unit V (volts)

Examples RES:HIGH 5

Query Syntax [SOURce:]RESistance:HIGH?

[SOURce:]RESistance:LOW?

Returned Parameters <NR3>


[SOURce:]RESistance:TRANsient:MODE

Channel Specific
This command selects the operating mode of the transient generator as follows in constant resistance

mode.

CONTinuous The transient generator puts out a continuous pulse stream upon receipt of a trigger.

PULSe

The transient generator puts out a single pulse upon receipt of a trigger.

TOGGle

The transient generator toggles between two levels upon receipt of a trigger.

Command Syntax [SOURce:]RESistance:TRANsient:MODE <mode>

Parameters CONTinuous | PULSe | TOGGle

*RST Value CONTinuous

Examples RES:TRAN:MODE PULS

Query Syntax [SOURce:]RESistance:TRANsient:MODE?

Returned Parameters <CRD>

Related Commands RES:TRAN:ALEV TRAN