Model 5861E

Section 2 Installation

Installation and Operation Manual

X-TMF-5861E-MFM-eng
Part Number: 541B107AAG
November, 2008

2-7 Electrical Interfacing

To insure proper operation the Model 5861E must be connected as shown in
Figure 2-2.

Note: There are several functions on the printed circuit board which are not
used by the Model 5861E. Avoid making connections marked in Figure 2-2
which are marked •. Also, the plug-in jumpers on printed circuit board are
not used by the Model 5861E and have no effect on performance.

The following minimum electrical connections must be made for new
installations:

Chassis Ground
0-5 Volt Signal Common
0-5 Volt Signal Output
+15 Vdc Supply
-15 Vdc Supply
Power Supply Common