Model 5861e – Brooks Instrument 5861E User Manual

Page 22

Advertising
background image

4-2

Model 5861E

Section 4 Maintenance

Installation and Operation Manual

X-TMF-5861E-MFM-eng

Part Number: 541B107AAG

November, 2008

Trouble

Possible Cause

Check/Corrective Action

Output stays at 0 Volts

Clogged Sensor.

Clean sensor. Refer to cleaning procedure. Section 4-4.

regardless of flow.

Output signal stays at +6.8V

Defective PC Board.

Replace PC Board. Refer to Section 4-4.

and there is no flow through
the meter.

Defective Sensor.

Replace Sensor Assembly. Refer to Section 4-4.

Meter grossly out of calibration. Partially clogged sensor.

Clean sensor, refer to the cleaning procedure.

Flow is higher than indicated.

Meter grossly out of calibration. Partially clogged restrictor.

Replace restrictor. Refer to Section 4-4.

Flow is lower than indicated.

Meter output oscillates.

Anticipate potentiometer out of adjustment. Adjust anticipate potentiometer. Refer to Section 3-4.

Faulty pressure regulator.

Check regulator output.

Defective PC Board.

Replace PC Board. Refer to Section 4-4.

Table 4-1 Bench Troubleshooting

Table 4-2 Sensor Troubleshooting