Selecting the measurement range, Calibrating the flow zero point – Burkert Type 8041 User Manual

Page 35

Advertising
background image

33

Adjustmentandcommissioning

Filter

Position of switch 3

slow
(Response time 10 to 90 % = 14 s)

OFF

fast
(Response time 10 to 90 % = 5 s)

ON

9.6.

selecting the measurement
range

The output signal is proportional to the measured flow velocity.
Switches 4 and 5 are used to adjust the measuring range of the
device to your application.

→

Position switches 4 and 5 to select the measuring range (see
Fig. 26, and the following table).

After the measuring range has been modified, the per-
centages set for the low and high switching thresholds are
applied to the new full scale selected.

Measuring range

Position of
switch 4

Position of switch
5

0 to 2 m/s

ON

OFF

0 to 5 m/s

OFF

ON

0 to 10 m/s

OFF

OFF

0 to calibrated full scale
(between 2 and 10 m/s)

ON

ON

9.7.

calibrating the flow zero point

→

Calibrate the device on commissioning and after each
maintenance task.

• Before calibrating the zero point on commissioning:

→

immerse the measuring element in the fluid for 24 h
before calibration.

• Before calibrating the zero point after each maintenance

task:

→

immerse the measuring element in the fluid for 1 h before
calibration.

→

Before calibration, ensure that the pipe does not contain
any air bubbles and that the fluid is not moving.

→

Fill the pipe with fluid.

→

Stop the flow.

→

Calibrate the "zero flow" point (see Fig. 28 and Fig. 29).

English

Type 8041