Burkert Type 8041 User Manual

Page 37

Advertising
background image

35

Adjustmentandcommissioning

A

→

Briefly press push button.

to calibrate automatically

→

Press and hold the push
button until...

to select the calibration of the
other parameter

B

→

Wait for 10 s.

to go back to the read mode

C

NO

Go back to Read mode
without confirming the
selection displayed.

Calibration in progress

Go back to Read mode.

.....

End of calibration: red LED
goes off?

YES

Zero point stored.

Red LED flashes briefly
twice: zero point has not
been calibrated.

→

Briefly press push button

Go back to Read mode
using the zero point of the
previous calibration.

→

Check that the calibration
instructions are respected
(see beginning chap. 9.7).

→

Restart calibration
of the zero flow
point.

...bargraph goes back to step 1, Fig.

28

Fig. 29: Calibration of the zero flow point, part 2

English

Type 8041