Display (English)
Display (German)
Description
Possible cause of error and
remedy
MAX
for temperature.
and the limit values if necessary.
Temperature < MIN Temperatur<MIN
The measured temperature is
below the set lower-range value
for temperature.
Product temperature is too low;
adjust the temperature range
and the limit values if necessary.
params inconsist
Parameter inkons.
Parameter is inconsistent.
Check the parameter settings.
The set parameters are contra-
dictory.
Example: Upper-range value,
pulse value and pulse duration
must be matched in such a way
that the combination fits for all
measured values.
missing EEPROM
ext EEPROM fehlt
The data memory module (DSB)
with the calibration data of the
sensor and the customer-specific
settings of the transmitter is not
plugged-in.
Insert the data storage module
(DSB/UMF33) in the correspond-
ing receptacle on the CPU
printed board UMC-30.
wrong EEPROM
falsches EEPROM
EEPROM of a former model (e.g.
UMC2 or UMF) has been
plugged-in as memory module.
internal communi-
cation faulty
interne Kommunika-
tion gestört
Communication between control
unit and transmitter is faulty.
Contact the device ven-
dor/customer service depart-
ment.
Information:
Error message: “Parameter is inconsistent” (system error 0x0400)?
To generate a list of the inconsistencies, first enter a valid password and then an inva-
lid password. The control unit will show a list of current errors (only once). The opera-
tor can then correct the inconsistent settings after entering a valid password.