3.4

Performance characteristics of the CMB sensor

3.4.1 Reference

conditions

•

Established flow profile

•

Inlet section has to correspond to mounting length

•

Operation is to be realized in the presence of downstream control valves

•

Measurement is to be realized in the absence of any gas bubbles

•

Flow tubes are to be kept clean at all times

•

Process temperature is to be regulated as specified in Section 3.6.1 Process temperature on
page 23

•

Process pressure is to be regulated as specified in Section 3.6.6 Process pressure range on
page 23

•

Ambient temperature is to range from + 10 °C to + 30 °C (50 °F to 86 °F)

•

Warm-up period: 15 minutes

•

Standard calibration is to be realized at 20 %, 50 % and 100 % (three times each)

•

High-frequency interference is to be regulated as specified in Section 17.2 Electromagnetic
compatibility on page 104

3.4.2

CMB flow ranges

Model

kg/h [lbs/min]

kg/h [lbs/min]

kg/h [lbs/min]

kg/h [lbs/min]

CMB-B

20 [0.7]

200 [7.3]

96 [3.5]

0.02 [0.001]

CMB-C

35 [1.3]

350 [12.9]

282 [10.4]

0.035 [0.00]

CMB-D

120 [4.4]

1,200 [44.1]

1,030 [37.8]

0.12 [0.00]

CMB-E

300 [11.0]

3,000 [110.2]

3,000 [110.2]

0.3 [0.0]

CMB-F

600 [22.0]

6,000 [220.5]

5,750 [211.3]

0.6 [0.0]

CMB-G

2,000 [73.5]

20,000 [734.9]

13,000 [477.7]

2 [0.1]

CMB-J

4,000 [147.0]

40,000 [1,469.7]

34,000 [1,249.3]

4 [0.1]

CMB-K

6,000 [220.5]

60,000 [2,204.6]

57,000 [2,094.4]

6 [0.2]

Zero point stability

(of range)

Max.

measuring range

Nominal

( p=1bar)

Min.

measuring range

Reference conditions: in conformity with IEC 770:
Temperature: 20 °C, relative humidity: 65 %, air pressure: 101.3 kPa
Fluid: water

CMB & CT OPERATING MANUAL
Part No 06EN003384 rev. 1.6

Page 15 of 106