11.5.2.2 Wiring diagram for the remote mount configuration of sensor and CT
Cable: Non-explosion proof applications
SLI2Y(ST)CY 5 x 2 x 0.5 mm² grey (max. 300 m)
Explosion-proof applications
SLI2Y(ST)CY 5 x 2 x 0.5 mm² blue (max. 300 m)
The outer shield is connected to the cable glands at both ends,
the inner shields are connected to each other and connected to the “Schirm / shield” terminal.
Schirm / shield
10
9
8
5
7
6
4
3
2
1
-
-
-
-
-
-
-
-
-
-
rosa / pink
grau / grey
schwarz / black
violett / violett
blau / blue
rot / red
gelb / yellow
grün / green
weiß / white
braun / brown
-
-
-
-
-
-
-
-
-
-
Schirm / shield
10
9
8
5
7
6
4
3
2
1
rosa / pink
grau / grey
schwarz / black
violett / violett
blau / blue
rot / red
gelb / yellow
grün / green
weiß / white
braun / brown
CMM, CMB, CMU with WAGO terminals
For terminal assignment, see Section 11.5.1 CT connections
Advices to cable glands: See also 10.1 “Installation conditions and cable glands” at page 35.