Density measurement, Accuracy – Fluid Components International CMB User Manual
Page 16

3.4.3 Density
measurement
The attainable accuracy depends on the selected calibration type.
Without calibration no density measurement is possible and the empty pipe recognition
is not available!
Model
without
3-Point
5-Point
CMB-B
CMB-C
CMB-D
5 g/l
3 g/l
CMB-E
5 g/l
3 g/l
CMB-F
5 g/l
3 g/l
CMB-G
5 g/l
3 g/l
CMB-J
5 g/l
3 g/l
CMB-K
5 g/l
3 g/l
Density accuracy
not available
no m
easurem
ent of d
ensity
3.4.4 Accuracy
Mass flow
Accuracy CMB-B to CMB-K
± 0.15% of actual flow + zero point stability (see Section 3.4.2 CMB
flow ranges)
Repeatability error
± 0.05% of actual flow (sensor with transmitter)
Additional measured values
Volume flow
± 0.2 % of actual value + zero point stability
Temperature
± 0.5 °C
Hysteresis n/a
Settling time
1 to 15 seconds
Startup drift
15 minutes
Long-term drift
± 0.02 % of upper-range value per year
Influence of ambient temperature
± 0.005 % per K
Influence of fluid temperature
Compensated
Influence of fluid pressure
For fluids: too small to be relevant