Sensor settings functional class, Sensor constant c – Fluid Components International CMB User Manual
Page 94

13.14 SENSOR SETTINGS functional class
The SENSOR SETTINGS functional class is comprised of the settings regarding the mass flow sensor.
SENSOR SETTINGS
SENSOR SETTINGS
Sensor
constant C
Sensor
constant C
Flow tube material
Flow tube material
SENSOR SETTINGS
functional class
Sensor constant
+0150.00
Flow tube
of [1.4571]
_______________
1.4571
1.4301
HC 4
HB 2
Tantal
Titan
Nickel
Special
Flow direction
Flow direction
Flow direction
[forward]
_______________
forward
reverse
forw. & revers
set by manufacturer
set by manufacturer
change only after re-
calibration (custody transfer)
13.14.1 Sensor constant C
Sensor constant C is the sensor calibration value for mass flow. This constant is defined when the flow-
meter is calibrated at the factory and can be found on the rating plate.
Sensor constant
+0150.00 kg/h
CAUTION:
Changing sensor constant C to a value that differs from the value on the rating plate
of the sensor connected to the flowmeter will result in false readings.
Page 94 of 106
CMB & CT OPERATING MANUAL