23ma max, 80ma max – Fluid Components International ST51 User Manual

Page 16

Advertising
background image

FLUID COMPONENTS INTERNATIONAL LLC

ST51 MASS FLOW

This page is subject to proprietary rights statement on last page

16

Doc. No. 06EN003389 Rev. -

Transmitter Circuit Calibration Check (Delta R Verification)

References

Delta ‘R’ Data Sheet

Equipment

FC88 Communicator or equivalent.©
DMV
Delta R Data Sheet-Match by serial numbers
2 Precision Decade Resistance boxes, 0.1% (Largest steps: 1K Ohm, smallest steps 0.01 Ohms)
Small flat Blade Screwdriver, 3/32 inches wide blade
FCI Normalization Cable, FCI part number 006407

Procedure

1. Verify all “D” mode calibration parameters are correct according to the meters Delta R Data Sheet, before starting verification.

2. Turn power off

3. Mark all sensor element wires connected to the circuit board, so they may be reconnected to the proper terminals. Disconnect the

wires.

4. Connect the resistance decade box to the electronics as per the appropriate diagram for the ST51.

NOTE:

Interconnector wiring (resistance decade box to electronics) must be 24 AWG and 45 inches long, to avoid any
inaccuracies in the Delta R verification, caused by improper wire lengths or wire gauges.

5. Set both decade boxes for the nominal resistance value (1000 ohms) +/- .01%

6. Connect DVM to the meters output termination and monitor the meter output.

7. Turn the power ON and allow the instrument 5 mins. To stabilize

8. With the FC88 connected, Press [T] [Enter] to view the Normal Operating Mode.

9. Adjust the Active Decade Box (Reference decade box remains fixed @ 1000 ohms) to achieve the appropriated Delta R for the

displayed flow value and output, noted on the meters Delta R Data Sheet.

10. Note the [C] mode and verify the meters displayed TCDR and REFR values corresponding to the displayed flow rate as per the

meters Delta R Data Sheet.

11. Return to the [T] mode to continue the verification.

REFERENCE

NORMALIZATION CABLE
FCI P/N 006407

BLACK

BLACK

ORANGE

ORANGE

1433-W DECADE RISISTOR GENERAL RADIO USA

98

76

X

G

23mA MAX

1K STEPS

H

L

10

K

2

0

1

4

5

3

ACTIVE

6

8

9 X

6 7

100 STEPS

80mA MAX

21

0

4

3

5

21

4

5

3

1433-W DECADE RISISTOR GENERAL RADIO USA

X

6 7 8 9

G

1K STEPS

23mA MAX

H

L

10

K

10

2

4

5

3

76

X98

6

100 STEPS

80mA MAX

0

1

2

5

3

4

1

2

4

5

3

Advertising