Section 12 default configuration, 1 user configuration table, Default configuration on – Hach-Lange POLYMETRON 9240 Operator Manual User Manual
Page 123

121
Section 12
Default Configuration
12.1 User configuration table
User Configuration
Option
Default Value
Possible Values
Selected Values
Date
MEASUREMENT
pH
11.20
10.00 - 12.00
Tgaz/Twater %
20
0000 - 0250
On-line measure time
10 minutes
0 - 9999 minutes
Smart rinsing
No
Yes or No
Max. rinse time
45 minutes
0 - 500 minutes
Cycle time
45 minutes
11 - 999 minutes
Search stability
No
Yes or No
Reactivation
frequency
24 hours
0 - 999 hours
Reactivation fixed date SMTWTFS 15:00:00
Any day and time
Graph time base
12 hours
4 hours
8 hours
12 hours
16 hours
20 hours
24 hours
ALARMS / ALARM 1
Mode
Limit
Limit
Active channel
Sample flow
None
Attrib
Channel 1
Channel 1 - 4
Limit
1.00 ppm
0 - 9990 ppm
Direction
Up
Up
Down
Delay
0 seconds
0 - 999 seconds
Hysteresis
0%
0 - 10%
Relay
N.O.
N.O.
N.C.
ALARMS / ALARM 2
Mode
Limit
Limit
Active channel
Sample flow
None
Attrib
Channel 1
Channel 1 - 4
Limit
1.00 ppm
0 - 9990 ppm
Direction
Up
Up
Down
Delay
0 seconds
0 - 999 seconds
Hysteresis
0%
0 - 10%
Relay
N.O.
N.O.
N.C.