Default configuration – Hach-Lange POLYMETRON 9240 Operator Manual User Manual

Page 124

Advertising
background image

122

Default Configuration

ALARMS / ALARM 3

Mode

Limit

Limit
Active channel
Sample flow
None

Attrib

Channel 1

Channel 1 - 4

Limit

1.00 ppm

0 - 9990 ppm

Direction

Up

Up
Down

Delay

0 seconds

0 - 999 seconds

Hysteresis

0%

0 - 10%

Relay

N.O.

N.O.
N.C.

ALARMS / ALARM 4

Mode

Limit

Limit
Active channel
Sample flow
None

Attrib

Channel 1

Channel 1 - 4

Limit

1.00 ppm

0 - 9990 ppm

Direction

Up

Up
Down

Delay

0 seconds

0 - 999 seconds

Hysteresis

0%

0 - 10%

Relay

N.O.

N.O.
N.C.

ALARMS / WARNING ALARM

Alarm

Yes

Yes
No

Accept

Manual

Auto.
Manual

Relay

N.O.

N.O.
N.C.

ALARMS / SYSTEM ALARM

Alarm

Yes

Yes
No

Relay

N.O.

N.O.
N.C.

mA OUTPUTS / OUTPUT 0

Type

0-20 mA

0-20 mA
4-20 mA

Mode

Linear

Linear
Dual
Logarithm

Low

0 ppb

0 - 9990 ppm

Middle

100 ppb

0 - 9990 ppm

High

1 ppm

0 - 9990 ppm

User Configuration

Option

Default Value

Possible Values

Selected Values