Section 8 calibration, 1 general, 2 calibration solution concentrations – Hach-Lange POLYMETRON 9240 Operator Manual User Manual

Page 95: 1 general 8.2 calibration solution concentrations, As the 92, Entitled

Advertising
background image

93

Section 8

Calibration

8.1 General

Before attempting to calibrate the analyzer, ensure that you have read and understood how to
enter and update data fields as described in the section entitled

Function keys on page 59

.

Note:

The instrument cannot be calibrated until at least one complete measurement cycle has been

successfully performed. An attempt to calibrate the instrument before this will result in a “Not authorized
message being displayed.

In order to obtain accurate measurement results, the analyzer should be calibrated on a regular
basis. Two basic methods of calibration are available:

Manual calibration

- The user can manually perform a calibration on an ad-hoc basis.

Automatic calibration

- Parameters can be set up so that the analyzer will perform a

calibration automatically on pre-defined days at a pre-defined time (see

Automatic calibration

setup on page 99

). This type of calibration is always done using known calibration solution

concentrations.

Note:

Before starting a calibration It is very important to ensure that the Reactivation Frequency

parameter (see

Reactivation frequency on page 82

) is greater than zero. If set to zero, the electrode

reactivation process will not take place during calibration and as such the measurement results may well be
out of limits.

To access the calibration menu, select CALIBRATION from the main menu and press the Enter
function key.

8.2 Calibration solution concentrations

Important Note:

The maximum concentration value for any calibration solution cannot be

greater than 2000 ppm

. Any value superior to this is outside the analyzer specifications and

cannot be processed.

Manual calibration

For a manual calibration, the calibration solutions must be between 100 ppb and 2000 ppm
and must respect the following rules:

LOW CAL SOL must be ≥ sample concentration of Na

+

and a minimum of 100 ppb

HIGH CAL SOL = (LOW CAL SOL x 10)

Example 1

: sample concentration of Na

+

= 20 ppb

LOW CAL SOL = minimum value = 100 ppb Na

+

HIGH CAL SOL = (100 ppb x 10) = 1000 ppb Na

+

Example 2

: sample concentration of Na

+

= 450 ppb

LOW CAL SOL = ≥ sample concentration of Na

+

= 500 ppb Na

+

HIGH CAL SOL = (500 ppb x 10) = 5000 ppb Na

+

Note:

The values LOW CAL SOL and HIGH CAL SOL must be entered into the system (see

One point

calibration on page 97

or

Two point calibration on page 98

).

Automatic calibration

For an automatic calibration, the calibration solution must be between 10 ppm and 2000 ppm
and must respect the following rules:

CAL SOL minimum = 10 ppm

CAL SOL = (sample concentration of Na

+

x 1000)

Example 1

: sample concentration of Na

+

= 0.1 ppb

CAL SOL = minimum value = 10 ppm Na

+

Example 2

: sample concentration of Na

+

= 1 ppm

CAL SOL = (1 ppm x 1000) = 1000 ppm Na

+

Note:

The value CAL SOL must be entered into the system (see

Automatic calibration setup on page 99

).

Advertising