87

User Setup

7.3.4 System and warning alarm table

The following table lists the different system and warning alarms:

Table 7 System and warning alarms

Message

Description

Category

Reset measure mod.

Reset the measure module

System

Reset LC module

Reset the local controller module

System

LC Error

Local controller module error

Warning

Err. measure mod.

Measure module error

Warning

Comms error

Communication error

Warning

Temp > 45°C (113°F)

Temperature too high

Warning

Temp < 5°C (45°F)

Temperature too low

Warning

Measure error

Measurement error

Warning

Meas.time over

Measurement cycle time too long

Warning

Cal.time over

Calibration cycle time too long

Warning

Flow x < min

Not enough sample

Warning

No condit.

No sample conditioner (< 50 mL)

System

Channel inhib.

The sample measurement is by-passed

Warning

Cal. sol < 100 mL

Calibration solution level low

Warning

Electro. < 10 mL

Electrolyte volume level low

Warning

Activ. sol. < 100 mL

Reactivation solution level low

Warning

Condit. < 100 mL

Conditioning solution level very low

Warning

P0 too high

Base measurement P0 is too high

Warning

P2 < 5 x P1

2nd point calibration < 5 times 1st point

Warning

Slope drift

Slope over tolerance

Warning

Offset drift

Offset over tolerance

Warning

Stabilize error

Stability criteria for accuracy not reached

Warning

Sample x < C

Sample concentration x is below threshold C

Warning

Sample x < 0 ppb

Sample concentration x is negative

Warning

Sample x > 1000 ppm

Sample concentration is too high

Warning