54

Installation

7.

For each channel, measure the pH in the conditioned sample to check if the pump ratios are
efficient enough to obtain a pH of around 11.0.

8.

If necessary, update the ratios of each channel to maintain a final constant pH of 11.0 ± 0.2.

4.4.10 Reference electrode installation

1.

Remove the reference
sensor from its box.

2.

Remove the plastic
reservoir from the bottom
(the storage solution is
KCl 3M) and install the
O-ring as shown.

3.

With care, turn the bottom
electrolyte tube ferrule
with a maximum ¼ turn to
lock it.

4.

Remove the plastic plug
on the entry port.