D A N G E R

Electrocution Hazard. Do not connect AC power to a 24 VDC powered model.

W A R N I N G

Potential Electrocution Hazard. A protective earth (PE) ground connection is required for both 100-240 VAC and
24 VDC wiring applications. Failure to connect a good PE ground connection can result in shock hazards and
poor performance due to electromagnetic interferences. ALWAYS connect a good PE ground to the controller
terminal.

N O T I C E

Install the device in a location and position that gives easy access to the disconnect device and its operation.

The controller can be purchased as either a 100-240 VAC powered model or a 24 VDC powered
model. Follow the appropriate wiring instructions for the purchased model.
The controller can be wired for line power by hard-wiring in conduit or wiring to a power cord.
Regardless of the wire used, the connections are made at the same terminals. A local disconnect
designed to meet local electrical code is required and must be identified for all types of installation. In
hard-wired applications, the power and safety ground service drops for the instrument must be 18 to
12 AWG.
Notes:

• The voltage barrier must be removed before making any electrical connections. After making all

connections, replace the voltage barrier before closing the controller cover.

• A sealing type cable gland and a power cord less than 3 meters (10 feet) in length with three 18-

gauge conductors (including a safety ground wire) can be used to maintain the NEMA
4X/IP66 environmental rating.

• Controllers can be ordered with AC power cords pre-installed. Additional power cords may also be

ordered.

• The DC power source that supplies power to the 24 VDC powered controller must maintain

voltage regulation within the specified 24 VDC-15% +20% voltage limits. The DC power source
must also provide adequate protection against surges and line transients.

Wiring procedure
Follow the numbered steps and

Table 1

or

Table 2

to wire the controller for power. Insert each wire

into the appropriate terminal until the insulation is seated against the connector with no bare wire
exposed. Tug gently after insertion to make a secure connection. Seal any unused openings in the
controller box with conduit opening sealing plugs.

Table 1 AC power wiring information (AC powered models only)

Terminal

Description

Color—N. America

Color—EU

1

Hot (L1)

Black

Brown

2

Neutral (L2)

White

Blue

Protective earth (PE) Ground lug

Green

Green and yellow

Table 2 DC power wiring information (DC powered models only)

Terminal

Description

Color—N. America

Color—EU

1

+24 VDC

Red

Red

2

24 VDC return

Black

Black

Protective earth (PE) Ground lug

Green

Green and yellow