Analog output connections

W A R N I N G

Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.

W A R N I N G

Potential electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the enclosure, use only
conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route cables in to the instrument.

Two isolated analog outputs are provided. Such outputs are commonly used for analog signaling or
to control other external devices. Make wiring connections to the controller as shown in

Figure 8

and

Table 3

.

Note:

Figure 8

shows the back of the controller cover and not the inside of the main controller compartment.

Figure 8 Analog output connections

Table 3 Output connections

Recorder wires

Circuit board position

Output 2-

4

Output 2+

3

Output 1-

2

Output 1+

1

1. Open the controller cover.
2. Feed the wires through the cable gland.
3. Adjust the wire as necessary and tighten the cable gland.
4. Make connections with twisted-pair shielded wire and connect the shield at the controlled

component end or at the control loop end.

• Do not connect the shield at both ends of the cable.
• Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher

than allowed.

• Maximum loop resistance is 500 ohm.

5. Close the controller cover and tighten the cover screws.
6. Configure outputs in the controller.

English 17