Discrete input wiring connections

W A R N I N G

Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.

W A R N I N G

Potential electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the enclosure, use only
conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route cables in to the instrument.

Three discrete inputs are provided for switch closure inputs or logic level voltage inputs. Make wiring
connections and configure jumper settings to the controller as shown in

Figure 9

,

Table 4

and

Figure 10

.

Note:

Figure 9

shows the back of the controller cover and not the inside of the main controller compartment.

Figure 9 Discrete input wiring connections

Table 4 Input connections

Discrete inputs

Connector position - Switch input

Connector position - Voltage

input

Input 1+

3

2

Input 1-

2

3

Input 2+

6

5

Input 2-

5

6

Input 3+

8

7

Input 3-

7

8

18 English