Figure 19 Feeder control function (phase low, overfeed timer)

1 Deadband

4 Time (x-axis)

7 Source (y-axis)

2 Setpoint

5 ON delay

3 Overfeed timer

6 OFF delay

Event control function (refer to

Figure 20

,

Figure 21

and

Figure 22

)

Option

Description

SET SETPOINT Sets the value where the relay will turn on.

DEADBAND

Sets a hysteresis so the relay will not swing unregulated when the process value
converges to the setpoint.

OnMax TIMER

Sets the maximum time the relay can stay on independent from the measured value
(default: + 0 min).

OffMax TIMER

Sets the maximum time the relay can stay off independent from the measured value
(default: + 0 min).

OnMin TIMER

Sets the minimum time the relay can stay on independent from the measured value
(default: + 0 min).

OffMin TIMER

Sets the minimum time the relay can stay off independent from the measured value
(default: + 0 min).