Hach-Lange POLYMETRON 9500 User Manual User Manual

Page 32

Advertising
background image

Option

Description

PWM CONTROL

Relay uses a PWM (Pulse Width Modulation) control depending on a
process value

FREQ CONTROL

Relay switches with a frequency depending on a process value

WARNING

Relay indicates warning and error conditions in sensors

6. From the RELAY SETUP menu, select SET TRANSFER and choose ACTIVE or INACTIVE.
7. From the RELAY SETUP menu, select FAIL SAFE and choose YES or NO.
8. From the RELAY SETUP menu, select ACTIVATION. The activation options for the selected

function appear.

9. Test the relay function to make sure it is properly energizing the connected device. To do relay

testing, Push the menu key and select TEST/MAINT>TEST RELAY.

Scheduler function (refer to

Figure 16

)

Option

Description

HOLD OUTPUTS

Holds outputs in the present ON or OFF state

RUN DAYS

Sets the weekday(s) that the relay operates.

START TIME

Sets the start time.

INTERVAL

Sets the time between activation cycles (default value: 5 min).

DURATION

Sets the period of time the relay is energized (default value: 30 sec).

OFF DELAY

Sets the time for additional hold/output time after the relay has been turned off.

Figure 16 Scheduler function

1 Duration

3 Interval

2 OFF delay

4 Time (x-axis)

Alarm function (refer to

Figure 17

)

Option

Description

LOW ALARM

Sets the value where the relay will turn on in response to decreasing measured value.
For example, if the low alarm is set for 1.0 and the measured value drops to 0.9, the
relay activates.

HIGH ALARM

Sets the value where the relay will turn on in response to increasing measured value.
For example, if the high alarm is set for 1.0 and the measured value increases to 1.1,
the relay activates.

30 English