Table 7 Sensor warning list

Warning

Definition

Resolution

EE SETUP ERR

Storage is corrupt. The values have been set
to the factory default.

Contact technical support.

EE RSRVD ERR

TEMP < 0 C

The process temperature is below 0 °C
(32 °F)

Increase the process temperature or stop
use until the process temperature is in the
sensor specification range.

TEMP > 50 C

The process temperature is above 50 °C
(120 °F)

Decrease the process temperature or stop
use until the process temperature is in the
sensor specification range.

RED AMPL LOW

Value falls below 0.03

Refer to

Table 6

on page 15 .

RED AMPL HIGH

Value is greater than 0.35

Call technical support.

BLUE AMPL LOW

Value is below 0.03

Refer to

Table 6

on page 15.

BLUE AMPL HIGH

Value is greater than 0.35

Call technical support.

CAP CODE FAULT The sensor cap code has become corrupt.

The code has been reset automatically to the
default cap and lot codes.

Complete the sensor setup cap procedure.
If no setup cap is available for the sensor
cap, call technical support.

Event list

The Event list keeps a log of changes to how data is recorded by the sensor. To show sensor events,
go to MENU>DIAGNOSTICS>[Select Sensor]>EVENT LIST.
Refer to

Table 8

.

Table 8 Event list for the sensor

Event

Description

ALT/PRESSURE UNIT CHANGE

Atmospheric pressure or altitude units have changed.

ALT/PRESSURE CHANGE

The value for altitude or atmospheric pressure has changed.

TEMP UNIT CHANGE

The units for temperature have changed.

MEAS UNIT CHANGE

A new unit of measurement has changed.

SALINITY CHANGE

The value for salinity has changed.

SET DEFAULT

Sensor settings have been reset to the default values.

SENSOR SETUP CHANGE

The sensor setup has changed.

CLEAN INTERVAL TIMER CHANGE

The time between sensor cleaning has changed.

SENSOR CAP LIFE TIMER CHANGE

The time between sensor cap replacements has changed.

16 English