Ldo sensor component list, Installation, Validate the sensor type – Hach-Lange LDO Sensor User Manual

Page 6

Advertising
background image

Figure 1 LDO sensor

1 Sensor cap

3 1-inch NPT

2 Temperature sensor

4 Connector, quick-connect (standard)

LDO Sensor component list

Make sure that all components shown in

Figure 2

have been received. If any items are missing or

damaged, contact the manufacturer or a sales representative immediately. Refer to

Figure 2

.

Figure 2 Sensor component list

1 LDO sensor

1

2 Calibration bags (2x)

1

Included user manual is not shown.

Installation

Validate the sensor type

D A N G E R

Explosion hazard. Connect only peripheral components that are clearly marked as certified for Class 1,
Division 2 Hazardous Locations.

N O T I C E

The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).

1. Go to the connector end of the cable.
2. Read the label on the connector end of the cable. For hazardous location certified sensors, the

label will show "Rated: Class 1 Division 2".

6 English